ZR

Zero Range

ZR


Command Group: Calibration

Access:

Network interfaces:

TAC protected

RW

CANopen

SDO2300sub11

MODbus

2220,2

ZR - Zero Range.

PROFIbus

read 3,17 write 4,17


Synopsis:
CommandResponse
GetZRR<Value>
SetZR <Value>OK (or ERR)
Valid forMin. valueMax. value
141.181 0 999999
141.184 0 999999
141.185 0 999999
141.186 0 999999
142.181 0 999999
142.184 0 999999
142.185 0 999999
142.186 0 999999
143.181 0 999999
143.184 0 999999
143.185 0 999999
151.181 0 999999
151.184 0 999999
151.186 0 999999
152.181 0 999999
152.184 0 999999
162.181 0 999999
179.181 0 999999
181.181 0 999999
64.181 0 999999
64.184 0 999999
69.181 0 999999
72.181 0 99999
78.183 0 99999



ZR Set the zero range - this is the range in increments within which the weighing scale can be zeroed. Issuing the ZR command without any parameter will return the current value. Set a new value by issuing the ZR command followed by the desired value, in the format ZR 5. Permitted values are between the lower limit of 0 and the upper limit of 99999.

Factory default setting: 0 (zero).
On LDM64, LDU78 and WCN122 A value of zero enables the previous zero range of +/-2% of max.
On other modules having the ZR command: A value of zero disables the RZ and SZ commands.

NOTE: Setting the zero range is protected by the CE sequence and the value is stored using the CS command.

See also RZ SZ CE, CS.