MR |
Multi range |
MR |
Command Group: Calibration | Access: | Network interfaces: | ||
TAC protected | RW | CANopen | SDO2300sub0D | |
MODbus | 2218,2 | |||
MR - Select multi range. | PROFIbus | read 3,13 write 4,13 |
|
|
MR Select multi range / multi interval.
0 = Multi interval, where the interval properties changes as the CM(n) limits are crossed.
1 = Multi range, where the highest entered range is kept until the weight reaches zero again.
Issuing the command without any parameters returns the current MR value.
To set a new value for MR, the command must be preceded by the CE XXXXX command, where XXXXX is the current TAC. The new MR value required is then input as a parameter of MR, in the format MR 1.
For further information, please see Calibration Procedure.
Factory default setting: 0.
MR is implemented in LDU78 from Firmware 78.183.v.2.47
The actual range can be shown using GG, GN, GL or GW. See OF.