MR

Multi range

MR


Command Group: Calibration

Access:

Network interfaces:

TAC protected

RW

CANopen

SDO2300sub0D

MODbus

2218,2

MR - Select multi range.

PROFIbus

read 3,13 write 4,13


Synopsis:
CommandResponse
GetMRM<Value>
SetMR <Value>OK (or ERR)
Valid forMin. valueMax. value
141.181 0 1
141.184 0 1
141.185 0 1
141.186 0 1
142.181 0 1
142.184 0 1
142.185 0 1
142.186 0 1
143.181 0 1
143.184 0 1
143.185 0 1
151.181 0 1
151.184 0 1
151.186 0 1
152.181 0 1
152.184 0 1
162.181 0 1
179.181 0 1
181.181 0 1
64.181 0 1
64.184 0 1
69.181 0 1
78.183 0 1



MR Select multi range / multi interval.

0 = Multi interval, where the interval properties changes as the CM(n) limits are crossed.
1 = Multi range, where the highest entered range is kept until the weight reaches zero again.

Issuing the command without any parameters returns the current MR value.
To set a new value for MR, the command must be preceded by the CE XXXXX command, where XXXXX is the current TAC. The new MR value required is then input as a parameter of MR, in the format MR 1. For further information, please see Calibration Procedure.
Factory default setting: 0.

MR is implemented in LDU78 from Firmware 78.183.v.2.47
The actual range can be shown using GG, GN, GL or GW. See OF.

See also CE, CM1, CM2, CM3, OF, GG, GN, GL, GW.