IZ

Instalation Zero.

IZ


Command Group: Calibration

Access:

Network interfaces:

TAC protected

W

CANopen

MODbus

IZ - Instalation Zero.

PROFIbus


Synopsis:
CommandResponse
SetIZOK (or ERR)
Valid forMin. valueMax. value
151.181    
152.181    
162.181    
179.181    
181.181    
64.181    
64.184    
78.183    



IZ Move the calibration zero point - this is used after instalation of an otherwise calibrated load cell. When installed, the instalation may add a bias to the load cell. This function can be used to move the CZ zeropoint without further calibration. IZ is subject to TAC control. The command returns ERR and has no action unless it is preceded by the CE XXXXX command, where XXXXX is the current TAC. Confirmation of action is provided by the return OK. For further information, please see Calibration Procedure.

See also CE, CZ, CG.