DP

Decimal position

DP


Command Group: Calibration

Access:

Network interfaces:

TAC protected

RW

CANopen

SDO2300sub0B

MODbus

2214,2

DP - Set the decimal point position

PROFIbus

read 3,11 write 4,11


Synopsis:
CommandResponse
GetDPP<Value>
SetDP <Value>OK (or ERR)
Valid forMin. valueMax. value
141.181 0 5
141.184 0 5
141.185 0 5
141.186 0 5
142.181 0 5
142.184 0 5
142.185 0 5
142.186 0 5
143.181 0 5
143.184 0 5
143.185 0 5
151.181 0 5
151.184 0 5
151.186 0 5
152.181 0 5
152.184 0 5
162.181 0 5
179.181 0 5
181.181 0 5
64.181 0 5
64.184 0 5
68.181 0 5
68.183 0 5
69.181 0 6
72.181 0 4
72.184 0 4
78.183 0 5
78.184 0 5
88.183 0 5
88.184 0 5
88.185 0 5



DP Set the decimal point position - this allows the decimal point to be positioned anywhere between leftmost and rightmost digits of the 5-digit output result. Permitted values are 0 for the rightmost position, and 5 for the leftmost position. To set a new value for DP, the command must be preceded by the CE XXXXX command, where XXXXX is the current TAC. The new DP value required is then input as a parameter of DP, in the format DP 2. For further information, please see Calibration Procedure. Factory default setting: 0.