CM3

Set Maximum3

CM3


Command Group: Calibration

Access:

Network interfaces:

TAC protected

RW

CANopen

SDO2300sub0F

MODbus

221C,2

CM3 - Set the maximum allowable output value #3.

PROFIbus

read 3,15 write 4,15


Synopsis:
CommandResponse
GetCM3M<Value>
SetCM3 <Value>OK (or ERR)
Valid forMin. valueMax. value
141.181 1 999999
141.184 1 999999
141.185 1 999999
141.186 1 999999
142.181 1 999999
142.184 1 999999
142.185 1 999999
142.186 1 999999
143.181 1 999999
143.184 1 999999
143.185 1 999999
151.181 1 999999
151.184 1 999999
151.186 1 999999
152.181 1 999999
152.184 1 999999
162.181 1 999999
179.181 1 999999
181.181 1 999999
64.181 1 999999
64.184 1 999999
69.181 1 999999
78.183 1 99999



CM3 - Maximum display value for range 3.

If CM3 = 0 then CM3 is disabled.
Set the maximum allowable output value in interval or range 3. Lower limit CM2+1, upper limit 999999.
Issuing the command without any parameters returns the current CM3 value.
This value will determine limit for range 3. When the weight becomes greater than CM3 then the output will change to ooooooo, signifying over-range.
To set a new value for CM3, the command must be preceded by the CE XXXXX command, where XXXXX is the current TAC. The new CM value required is then input as a parameter of CM, in the format CM 3 4010. For further information, please see Calibration Procedure.

Factory default setting: Max 1 = 999999, Max 2 = 0, Max 3 = 0.

CM3 is implemented in LDU78 from Firmware 78.183.v.2.47

See also CE, CI, DS, DP, CZ, CG, ZT, AZ, CS, FD, UR, MR, CM1, CM2.