CM3 |
Set Maximum3 |
CM3 |
Command Group: Calibration | Access: | Network interfaces: | ||
TAC protected | RW | CANopen | SDO2300sub0F | |
MODbus | 221C,2 | |||
CM3 - Set the maximum allowable output value #3. | PROFIbus | read 3,15 write 4,15 |
|
|
CM3 - Maximum display value for range 3.
If CM3 = 0 then CM3 is disabled.
Set the maximum allowable output value in interval or range 3.
Lower limit CM2+1, upper limit 999999.
Issuing the command without any parameters returns the current CM3 value.
This value will determine limit for range 3. When the weight becomes greater than CM3 then the output will change to ooooooo,
signifying over-range.
To set a new value for CM3,
the command must be preceded by the CE XXXXX command, where XXXXX is the current TAC.
The new CM value required is then input as a parameter of CM, in the format CM 3 4010.
For further information, please see Calibration Procedure.
Factory default setting: Max 1 = 999999, Max 2 = 0, Max 3 = 0.
CM3 is implemented in LDU78 from Firmware 78.183.v.2.47
See also CE, CI, DS, DP, CZ, CG, ZT, AZ, CS, FD, UR, MR, CM1, CM2.