CE

Calibrate enable

CE


Command Group: Calibration

Access:

Network interfaces:

RW

CANopen

SDO2300sub03

MODbus

2204,2

CE - Set the calibration functions to the enabled state

PROFIbus

read 3,3 write 4,3


Synopsis:
CommandResponse
GetCEE+<Value>
SetCE <Value>OK (or ERR)
Valid forMin. valueMax. value
141.181 0 65535
141.184 0 65535
141.185 0 65535
141.186 0 65535
142.181 0 65535
142.184 0 65535
142.185 0 65535
142.186 0 65535
143.181 0 65535
143.184 0 65535
143.185 0 65535
151.181 0 65535
151.184 0 65535
151.186 0 65535
152.181 0 65535
152.184 0 65535
162.181 0 65535
179.181 0 65535
181.181 0 65535
64.181 0 65535
64.184 0 65535
68.181 0 65535
68.183 0 65535
69.181 0 65535
72.181 0 65535
72.184 0 65535
78.183 0 65535
78.184 0 65535
88.183 0 65535
88.184 0 65535
88.185 0 65535



CE Set the calibration functions to the enabled state. This command must be issued PRIOR to any attempt to set the calibration parameters CM, CI, MR, DS, DP, CZ, CG, ZT, ZR, FD, LC, LN or CS. Issuing the command without any parameters results in the response E+XXXXX where XXXXX is the Traceable Access Code (TAC). This is an internal code that is used to record any changes in the calibration settings of the device. This is a critical feature that is required for “approved” applications, as it provides for the control of access to any command that has the potential of changing the weigher calibration value.

See also CE, CI, DS, DP, CZ, CG, ZT, AZ, CS, FD, UR,