AG

Absolute Gain

AG


Command Group: Calibration

Access:

Network interfaces:

TAC protected

RW

CANopen

SDO2300sub01

MODbus

2200,2

AG - Set the absolute gain point

PROFIbus

read 3,1 write 4,1


Synopsis:
CommandResponse
GetAGG<Value>
SetAG <Value>OK (or ERR)
Valid forMin. valueMax. value
141.181 -33000 33000
141.184 -33000 33000
141.185 -33000 33000
143.181 -33000 33000
143.184 -33000 33000
143.185 -33000 33000
151.181 -33000 33000
152.181 -33000 33000
162.181 -33000 33000
179.181 -33000 33000
181.181 -33000 33000
64.181 -33000 33000
64.184 -33000 33000
68.181 -22000 22000
72.181 -32000 32000
72.184 -32000 32000
78.183 -22000 22000
88.183 -22000 22000
88.184 -22000 22000
88.185 -22000 22000



AG The command AG is used as absolute gain (or measuring range) for all weight calculations and will setup in mV/V.
Permitted values are ± 22000 (= ± 2.2000 mV/V) for LDU68, LDU78 and LDM88.
Permitted values are ± 33000 (= ± 3.3000 mV/V) for all others.

Factory default: 20000 @ 2.0000 mV/V input signal.

The serial interface offers two parameters: “mV/V” and “value”.

When calibration is done over the network the second value is taken from the CV register,
except System80/MCS-64 Gateways (for LDM88) that takes the second parameter from the CG value (normally 20000).

See also CE, CM, CI, DS, DP, CZ, CG, ZT, AZ, CS, FD, UR,