ZT

Zero track

ZT


Command Group: Calibration

Access:

Network interfaces:

TAC protected

RW

CANopen

SDO2100sub12

MODbus

2122,2

ZT - enables or disables the zero tracking

PROFIbus

read 18,18 write 31,18


Synopsis:
CommandResponse
GetZTZ:<Value>
SetZT <Value>OK (or ERR)
Valid forMin. valueMax. value
141.181 0 255
141.184 0 255
141.185 0 255
141.186 0 255
142.181 0 255
142.184 0 255
142.185 0 255
142.186 0 255
143.181 0 255
143.184 0 255
143.185 0 255
151.181 0 255
151.184 0 255
151.186 0 255
152.181 0 255
152.184 0 255
162.181 0 255
179.181 0 255
181.181 0 255
64.181 0 255
64.184 0 255
68.181 0 255
68.183 0 255
69.181 0 65535
72.181 0 255
72.184 0 255
78.183 0 255
78.184 0 1
88.183 0 255
88.184 0 255
88.185 0 255



ZT Zero tracking - this command enables or disables the zero tracking. Parameter = 0 disables the zero tracking and parameter = 1 enables the zero tracking. Issuing the command without any parameters returns the current ZT value. The command returns ERR and has no updating action unless it is preceded by the CE XXXXX command, where XXXXX is the current TAC. Zero tracking will be performed only on results less than +/-0.5d at a rate of 0.4 d/sec where d = display step size (see DS command). The zero can only be tracked to +/- 2% of max (see CM command). Factory default setting: 0.